Universal 320
Benchtop Centrifuge
Non-Refrigerated/Refrigerated
|
|
|
OPERATING AND FEATURES
- Accommodates various tube systems from 0.2ml to 100ml, hematocrit capillaries as well as microtitre plates, culture plates
- Microprocessor-controlled
- 10 complete running programs can be stored and recalled
- Run-up/run-down times: Input of acceleration & braking ramps in levels 1-9
- Easy to read; digital display
- Speed RPM: Input in increments of 10
- Speed RCF: Input in increments of 10
- Running Time: Input in minutes and seconds, max 99 minutes 59 seconds
- Impulse key - short cycle mode
- Radius (r): Input in mm when RPM or RCF need to be determined
- During centrifugation the actual values of the parameters are displayed
- Temperature, T/C°:With the Universal 320R input in increments of 1°C within a range of -20°C to + 40°C
- Fast Cool Function on Universal 320R
- Brushless frequency drive
SAFETY
- Metal housing and metal lid
- Stainless steel centrifuging chamber of stainless steel, easy to clean while providing additional safety
- Automatic lid-locking
- Lid-locking and lid holding during rotor run
- Emergency lid lock release
- Motor overheating protection
- Chamber overheating protection with Universal 320R
- Imbalance switch-off
- Automatic rotor recognition
- Lid dropping protection
SPEED
- Max. Speed with angle rotor 15,000 RPM, 21,382 RCF
- Max. Speed with swing-out rotor 5,000 RPM, 4,193 RCF
- Max. Speed with hematocrit rotor 15,000 RPM, 21,382 RCF
- Max. Speed with hematocrit rotor 4,000 RPM, 2,093 RCF
| Model
| Weight (Kg)
| Dimensions W x H x D (mm)
| Available Voltage
| Part Number
|
| Universal 320
| 29
| 346 X 395 X 520
|
|
|
| Universal 320R
| 52
| 346 X 401 X 695
|
| |
| Rotor Number
| Rotor Type
| Places
| Tube Size (mL)
| Max RCF
|
|
1624
| SWING-OUT
| 4
| 1 - 100 ML & CYTO
| 2,665
|
1611
| SWING-OUT
| 8
| 1.6 - 15 ML
| 2,415
|
1628
| SWING-OUT
| 12
| 1.6 - 15 ML
| 3,396
|
1617
| SWING-OUT
| 8
| 15 & 50 ML
| 3,857
|
1619
| SWING-OUT
| 6
| 15 & 50 ML
| 2,701
|
1622
| SWING-OUT
| 2
| 4 MICROTITRATION PLATES
| 2,093
|
1650
| HEMATOCRIT
| 24
| N/A
| 21,382 / 7,697
|
1626
| CYTO
| 6
| N/A
| 2,039
|
1648
| CYTO
| 8
| N/A
| 2,218
|
1620A
| ANGLE
| 6
| 1.5 - 85 ML
| 9,418
|
1613
| ANGLE
| 12
| 5 - 15 ML
| 4,105
|
1615
| ANGLE
| 12
| 5 & 15 ML
| 16,421 / 9,237
|
1612
| ANGLE
| 12
| 0.4 - 2.0 ML
| 23,907
|
1645
| SWING-OUT
| 4
| MICROTITRATION PLATES, CULTURE PLATES
| 1,896
|
1689
| ANGLE
| 30
| 0.4 - 2.0 ML
| 21,255
|
1494
| SWING-OUT
| 4
| 1.5 - 100 ML
| 4,193
|
1420
| ANGLE
| 24
| 0.4 - 2.0 ML
| 21,382
| | | | | | | | | | | | | | | | |
|
**Additional Options**
P/N 1406-50, Universal 320R
• Centrifuge may be heated or cooled within a range of 0°C to + 90°C
• Customer must provide connection to an external refrigerated/heating circulator.
Exact temperature range dependent on circular performance
P/N 1406-20, Universal 320R
• Centrifuge provided with connection for flushing with nitrogen
• Without monitoring and controlling system
|
|